DocumentCode :
3465154
Title :
A Novel Calibration Method for Noninvasive Blood Pressure Measurement Using Pulse Transit Time
Author :
Yan, Y.S. ; Zhang, Y.T.
Author_Institution :
Chinese Univ. of Hong Kong, Shatin
fYear :
2007
fDate :
19-22 Aug. 2007
Firstpage :
22
Lastpage :
24
Abstract :
This paper proposes a new calibration method for arterial blood pressure (BP) estimation using pulse transit time (PTT). Different from the traditional calibration methods, this new approach features a fast and convenient procedure by modeling the effects on distal arteries of external pressure applied by upper arm cuff. Two rounds of experiments (about two months apart) were conducted on eleven healthy subjects. The results show that the obtained calibration curves can predict the mean BP within 4.5plusmn6.6mmHg of the reference for all the tested subjects during steady state and immediately after exercise. It is also found that the new method has good repeatability within two months. The preliminary results indicate that the new approach has the potential to be used for noninvasive and cuffless measurement of arterial BP in wearable devices.
Keywords :
biomedical equipment; blood pressure measurement; calibration; arterial blood pressure estimation; calibration method; noninvasive blood pressure measurement; pulse transit time; wearable medical device; Arteries; Biomedical measurements; Biomedical monitoring; Blood pressure; Calibration; Fingers; Heart; Pressure measurement; Pulse measurements; Time measurement; Blood pressure; hydrostatic pressure; mobile health; pulse transit time; wearable medical device;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Medical Devices and Biosensors, 2007. ISSS-MDBS 2007. 4th IEEE/EMBS International Summer School and Symposium on
Conference_Location :
Cambridge
Print_ISBN :
978-1-4244-1346-1
Electronic_ISBN :
978-1-4244-1346-1
Type :
conf
DOI :
10.1109/ISSMDBS.2007.4338283
Filename :
4338283
Link To Document :