    Quality control system for biomedical instrumentation

  • Author

    Jobbagy, A. ; Pataricza, Andras

  • Author_Institution
    Dept. of Meas & Instrum. Eng., Tech. Univ., Budapest, Hungary
  • fYear
    1988
  • fDate
    4-7 Nov. 1988
  • Firstpage
    1233
  • Abstract
    The authors describe the self-test and final test of a three-channel ECG device. It is also shown how the self-test routines are used for calibration, final test, and field service. In order to promote testing, an almost hardware-level controllability is assured through the remote-control connector. Though the instrument has the usual built-in calibration signal, an extra calibration signal is available which serves for testing the analog channels together with the electrode cables during service.<>
  • Keywords
    biomedical equipment; quality control; 3-channel ECG device; analog channels; biomedical instrumentation; built-in calibration signal; electrode cables; field service; final test; hardware-level controllability; quality control system; remote-control connector; self-test routines; testing;