• DocumentCode
    347080
  • Title

    Spatial sampling of neocortical activity by electrode grids

  • Author

    Hoekema, Rudi ; Huiskamp, Geertjan ; Wieneke, George

  • Author_Institution
    Dept. of Clinical Neurophysiol., Univ. Hosp. Utrecht, Netherlands
  • Volume
    1
  • fYear
    1999
  • fDate
    1999
  • Abstract
    In epilepsy surgery, electrode grids are often used to measure electrocorticograms (ECoGs). In these electrode grids, the distance between contacts is usually 10 mm. This may not be adequate to sample neocortical electrical activity. In this study, ECoGs were simulated, assuming the location of the electrical sources in the brain generating epileptic spikes. It is concluded that for adequately measuring the electric field generated by cortical sources, the contact distance should be decreased to 5 mm
  • Keywords
    biomedical electrodes; electroencephalography; surgery; 10 mm; 5 mm; ECoGs; contact distance; electrocorticograms measurement; electrode grids; epilepsy surgery; epileptic spikes generation; neocortical electrical activity sampling; Conducting materials; Conductors; Contacts; Electric variables measurement; Electrodes; Epilepsy; Hospitals; Neurophysiology; Sampling methods; Surgery;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    [Engineering in Medicine and Biology, 1999. 21st Annual Conference and the 1999 Annual Fall Meetring of the Biomedical Engineering Society] BMES/EMBS Conference, 1999. Proceedings of the First Joint
  • Conference_Location
    Atlanta, GA
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-5674-8
  • Type

    conf

  • DOI
    10.1109/IEMBS.1999.802530
  • Filename
    802530