DocumentCode :
3470931
Title :
Lithium Fluoride TLD dose quality
Author :
Ormond, Eugene C. ; Bozman, Dan R. ; Cordova, Steven R. ; Mitchell, David R. ; Oliver, Bryan V. ; Good, Douglas E. ; Henderson, D.J. ; Hogge, Keith ; Howe, Russell A. ; Huber, Stephen R. ; Larsen, M.L. ; Mitchell, Stephen E. ; Mitton, Charles V. ; Molina,
Author_Institution :
Sandia Nat. Labs. Albuquerque, Albuquerque, NM, USA
fYear :
2013
fDate :
16-21 June 2013
Firstpage :
1
Lastpage :
5
Abstract :
Lithium Fluoride (LiF) Thermoluminescent Dosimeters (TLD´s) have been used extensively as a diagnostic in radiographic facilities. For over three years we have used precision-calibrated TLD´s in our facility. Said facility consists of two radiographic sources, each with a dose rating of 4-Rad at 1 m. The calibration and fielding of TLD´s will be examined for accuracy and long term stability. TLD´s will be evaluated in single point measurements and multi-point arrays. Overlaying three-chip TLD measurements will be compared to single chip TLD measurements. TLD results will also be compared to Pin Diodes for routine shot diagnosis and evaluation of shielding around the sources.
Keywords :
calibration; lithium compounds; p-i-n diodes; shielding; thermoluminescent dosimeters; LiF; lithium fluoride TLD dose quality; long term stability; multipoint array; pin diode; radiographic source; thermoluminescent dosimeter; three-chip TLD measurement; Accuracy; Calibration; Hardware; Lithium; PIN photodiodes; Semiconductor device measurement; Standards;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Pulsed Power Conference (PPC), 2013 19th IEEE
Conference_Location :
San Francisco, CA
ISSN :
2158-4915
Type :
conf
DOI :
10.1109/PPC.2013.6627700
Filename :
6627700
Link To Document :
بازگشت