DocumentCode
3471255
Title
AC-DC difference measurements relative to Josephson generated voltages
Author
Palafox, L. ; Behr, R. ; Funck, T.
Author_Institution
Phys.-Tech. Bundesanstalt, Braunschweig, Germany
fYear
2010
fDate
13-18 June 2010
Firstpage
91
Lastpage
92
Abstract
In an extension from their original use as references for measuring dc voltage, Josephson generated voltages from programmable arrays have been employed to measure waveforms differentially. In this paper, the alternative of using these Josephson voltages as a reference for ac-dc difference measurements using thermal converters and sampling techniques is presented.
Keywords
AC-DC power convertors; Josephson effect; voltage measurement; AC-DC difference measurements; Josephson generated voltages; programmable arrays; sampling techniques; thermal converters; Analog-digital conversion; Current measurement; DC generators; Electromagnetic measurements; Frequency synchronization; Measurement uncertainty; Sampling methods; Synthesizers; Voltage measurement; Voltmeters;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM), 2010 Conference on
Conference_Location
Daejeon
Print_ISBN
978-1-4244-6795-2
Type
conf
DOI
10.1109/CPEM.2010.5543958
Filename
5543958
Link To Document