High-resolution absolute SPECT quantitation for I-131 distributions used in the treatment of lymphoma: a phantom study
Author
Trotter, D. E Gonzalez ; Jaszczak, R.J. ; Bowsher, J.E. ; Akabani, G. ; Greer, K.L.
Author_Institution
Dept. of Radiol., Duke Univ. Med. Center, Durham, NC, USA
Volume
3
fYear
2000
fDate
2000
Abstract
Clinical lymphoma I-131 radioimmunotherapy (RIT) studies are underway at the authors´ institution. The effectiveness of treatment can be evaluated through accurate time dependent three-dimensional dosimetry calculations based on SPECT images. The authors compared the absolute quantitative SPECT performance of a high-resolution rotating parallel-hole collimator (RPHC) designed to minimize septal penetration with that of a traditional medium-energy parallel-hole collimator (MEPHC). They scanned a phantom consisting of 4 spheres having inner diameters of 2.2 and 2.8 cm and whose activity concentration ratios with respect to the surrounding medium were between 5:1 and 17:1. Images were reconstructed using the OSEM algorithm (10 subsets, 20 iterations) with spatial resolution modeling. Quantitation of the MEPHC SPECT images yielded a large overestimation of the surrounding activity concentration (50%) due to septal penetration. Quantitation of the RPHC SPECT image yielded a surrounding medium activity concentration within 8% of the expected value, and the activity within the spheres was estimated to within 20% of their absolute expected values, and 12% of their relative expected values. It is concluded that for the present quantitation application the benefits of septal penetration suppression by the RPHC outweigh the increased image noise due to low count rate. These results point to the feasibility of using a large field-of-view (FOV) collimator based on the imaging principles of the RPHC for accurate clinical SPECT quantitation of I-131 RIT
Keywords
cancer; image reconstruction; image resolution; iterative methods; medical image processing; radiation therapy; single photon emission computed tomography; 2.2 to 2.8 cm; I; I-131 distributions; absolute quantitative SPECT performance; activity concentration ratios; clinical lymphoma I-131 radioimmunotherapy studies; high-resolution absolute SPECT quantitation; high-resolution rotating parallel-hole collimator; lymphoma treatment; medical diagnostic imaging; nuclear medicine; septal penetration suppression; spatial resolution modeling; traditional medium-energy parallel-hole collimator; Biomedical imaging; Collimators; Dosimetry; Imaging phantoms; Legged locomotion; Neoplasms; Radioactive materials; Radiology; Reconstruction algorithms; Spatial resolution;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2000 IEEE
Conference_Location
Lyon
ISSN
1082-3654
Print_ISBN
0-7803-6503-8
Type
conf
DOI
10.1109/NSSMIC.2000.949214
Filename
949214
Link To Document