  • Abstract
    The problems of the reconstruction of the trajectory of γ-rays and of the identification of their interaction points in segmented HpGe detectors are discussed in view of their importance in the design of the next generation of detector arrays for γ-ray spectroscopy studies. In fact, the level of tracking of the γ-ray interactions in the medium controls the overall detection efficiency. Tracking algorithms taking into accounts the detection segmentation together with the pulse shape analysis have been developed making use of the GEANT simulation package. The present work focuses on two main points. One is how to disentangle the pulse shape analysis effects and investigate their possibilities, the other is the study of the consequences that different types of segmentation have on the overall efficiency. Two geometries have been chosen, that of a single cylindrical detector and a spherical shell
  • Keywords
    gamma-ray detection; gamma-ray spectrometers; germanium radiation detectors; γ-ray detectors; γ-ray spectroscopy; GEANT; Ge; HPGe detectors; detection efficiency; pulse shape analysis; trajectory; Algorithm design and analysis; Analytical models; Gamma ray detection; Gamma ray detectors; Geometry; Packaging; Pulse shaping methods; Sensor arrays; Shape; Spectroscopy;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2000 IEEE
  • Conference_Location
    Lyon
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-6503-8
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2000.949295
  • Filename
    949295