DocumentCode :
3473134
Title :
A physics based model of pixellated semiconductor gamma camera
Author :
Glière, A. ; Koenig, A. ; Mathy, F. ; Hugonnar, P.
Author_Institution :
LETI-CEA Technol. Avancees, CEA, Centre d´´Etudes Nucleaires de Grenoble, France
Volume :
3
fYear :
2000
fDate :
2000
Abstract :
A solid state CdZnTe gamma camera simulator, taking into account the gamma-ray emission and attenuation in the examined phantom, a geometric collimator model and the CdZnTe semiconductor detector response, has been developed. In a first step, the energy dependent spectrometric response of the planar semiconductor pixels is computed from first physical principles (gamma-ray interaction with the crystal and charge collection in the semiconductor detector), as well as from specific acquisition parameters (front-end electronics signal filtering, pulses classification in a pulse height versus rise time biparametric spectrum and bi-dimensional window based pulses selection). In a second step, the uncollided and once scattered photon fluxes, incident to each gamma camera´s pixel, are computed by the means of ray tracing through the phantom´s CAD model. They are then combined with a geometric collimator model. The energy dependent pixel response is finally used to generate the simulated image. Each component of the model has been compared to experimental data. Finally, a correct qualitative agreement was found between simulated images and measurements performed on a prototype platform
Keywords :
biomedical equipment; cameras; gamma-ray detection; radioisotope imaging; semiconductor counters; semiconductor device models; CdZnTe; bi-dimensional window based pulses selection; energy dependent pixel response; energy dependent spectrometric response; gamma-ray crystal interaction; geometric collimator model; medical diagnostic imaging; medical instrumentation; nuclear medicine; physics based model; pixellated semiconductor gamma camera; planar semiconductor pixels; ray tracing; rise time biparametric spectrum; simulated image generation; solid state CdZnTe gamma camera simulator; Attenuation; Cameras; Collimators; Computational modeling; Gamma ray detection; Gamma ray detectors; Imaging phantoms; Physics; Solid modeling; Solid state circuits;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2000 IEEE
Conference_Location :
Lyon
ISSN :
1082-3654
Print_ISBN :
0-7803-6503-8
Type :
conf
DOI :
10.1109/NSSMIC.2000.949298
Filename :
949298
Link To Document :
بازگشت