DC performance of a high accuracy thermal transfer standard
Author
Early, M.D. ; Garcocz, M. ; Christian, L.A.
Author_Institution
Meas. Stand. Lab. of New Zealand (MSL), Lower Hutt, New Zealand
fYear
2010
fDate
13-18 June 2010
Firstpage
201
Lastpage
202
Abstract
The dc performance of a widely-used high-accuracy commercial thermal transfer standard is found to be significantly affected by a variable contact resistance on its lower impedance ranges. This has a direct impact on the AC voltages measured by the instrument on these ranges. Cleaning the contacts and ensuring clearance around an input common mode choke effectively resolves the problem.
Keywords
cleaning; contact resistance; digital voltmeters; inductors; temperature measurement; AC voltages; calibrated precision DC DVM; contact cleaning; high-accuracy commercial thermal transfer standard; input common mode choke; variable contact resistance; Calibration; Cleaning; Electrical resistance measurement; Electromagnetic measurements; Frequency; Impedance; Instruments; Laboratories; Measurement standards; Voltage;