• DocumentCode
    3474104
  • Title

    Design considerations for small animal PET devices with mm resolution

  • Author

    Correia, J.A. ; Burnham, C.A. ; Kaufman, D. ; Fischman, A.J.

  • Author_Institution
    Massachusetts Gen. Hospital, Boston, MA, USA
  • Volume
    3
  • fYear
    2000
  • fDate
    2000
  • Abstract
    The authors have recently reported a single-plane LSO-based-PET instrument for imaging small animals. The thin detector employed 360 one mm LSO crystals. Images obtained were characterized by resolution approaching the sampling limit of 1 mm. Work directed at extending the performance of this instrument is reported here. A second-generation, single plane device is under construction and preliminary results for a multiplanar instrument have also been obtained
  • Keywords
    biological techniques; image resolution; positron emission tomography; 1 mm; LSO crystals; biological research instrumentation; multiplanar instrument; nuclear medicine; resolution approaching sampling limit; second-generation single plane device; single-plane LSO-based-PET instrument; small animals imaging; thin detector; Animals; Crystals; Detectors; Energy resolution; Image resolution; Instruments; Positron emission tomography; Prototypes; Signal resolution; Torso;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2000 IEEE
  • Conference_Location
    Lyon
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-6503-8
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2000.949344
  • Filename
    949344