DocumentCode
3474746
Title
Performance measurement of a new high resolution detector system for I-131 thyroid studies
Author
Bussmann, N. ; Engels, R. ; Fuss, L. ; Kemmerling, G. ; Reinartz, R. ; Langen, K.-J. ; Schelten, J. ; Ziemons, K.
Author_Institution
Zentrallabor fur Elektronik, Forschungszentrum Julich GmbH, Germany
Volume
3
fYear
2000
fDate
2000
Abstract
A 2-dimensional detector system for high resolution thyroid I-131 scintigraphy was developed. It has a sensitive area of 4 cm×4 cm and consists of a lead-collimator and an array of 10×10 EGO crystals combined with a position sensitive photomultiplier. The spatial resolution and the sensitivity of the detector has been measured and compared to two commercially available gamma-cameras. Furthermore first patient measurements have been carried out
Keywords
biological organs; biomedical equipment; image resolution; photomultipliers; radioisotope imaging; 2-dimensional detector system; EGO crystals; I; detector sensitivity; high resolution thyroid I-131 scintigraphy; medical diagnostic imaging; medical instrumentation; nuclear medicine; position sensitive photomultiplier; spatial resolution; Cancer; Collimators; Crystals; Gamma ray detection; Gamma ray detectors; Head; Imaging phantoms; Measurement; Photomultipliers; Spatial resolution;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2000 IEEE
Conference_Location
Lyon
ISSN
1082-3654
Print_ISBN
0-7803-6503-8
Type
conf
DOI
10.1109/NSSMIC.2000.949375
Filename
949375
Link To Document