DocumentCode
3475058
Title
Direct effective efficiency measurement for thermoelectric power sensors with true-twin microcalorimeter by a switching method
Author
Vremera, E. ; Brunetti, L.
Author_Institution
Tech. Univ. of Iasi, Iasi, Romania
fYear
2010
fDate
13-18 June 2010
Firstpage
702
Lastpage
703
Abstract
The paper describes a new power sensors´ calibration method suitable for the twin type coaxial microcalorimeter. This method allows an intrinsic compensation of the lines losses, thus producing direct results, as in an ideal unitary-gain microcalorimeter. Furthermore, it does not require the microcalorimeter calibration step with the short circuit and an equation similar to that for the ideal system is applied for the measurand computation.
Keywords
calibration; calorimeters; coaxial cables; microwave measurement; power measurement; sensors; switching; thermoelectric power; coaxial microcalorimeter; direct effective efficiency measurement; line losses; microcalorimeter calibration; power sensor calibration method; short circuit; switching method; thermoelectric power sensor; true-twin microcalorimeter; unitary-gain microcalorimeter; Calibration; Coaxial components; Dielectric measurements; Electromagnetic measurements; Equations; Hardware; Power measurement; Sensor phenomena and characterization; Thermal sensors; Thermoelectricity;
fLanguage
English
Publisher
ieee
Conference_Titel
Precision Electromagnetic Measurements (CPEM), 2010 Conference on
Conference_Location
Daejeon
Print_ISBN
978-1-4244-6795-2
Type
conf
DOI
10.1109/CPEM.2010.5544182
Filename
5544182
Link To Document