Abstract
This paper describes a high-resolution and high-precision autotransformer ratio capacitance bridge that has been built and tested at PTB for dielectric-constant gas thermometry (DCGT). Its main component is a home-made high-precision 1:1 inductive voltage divider (IVD) used in a autotransformer configuration. Special emphasis is given to the estimation of the uncertainty of the determination of small capacitance changes considering correlations.
Keywords
autotransformers; bridge circuits; permittivity; thermometers; voltage dividers; dielectric-constant gas thermometry; high-precision autotransformer ratio capacitance bridge; high-resolution autotransformer ratio capacitance bridge; home-made high-precision 1:1 inductive voltage divider; Bridge circuits; Capacitance measurement; Dielectric constant; Dielectric measurements; Equations; Helium; Kelvin; Magnetic cores; Temperature; Uncertainty;