• DocumentCode
    3477096
  • Title

    Microwire-wound resistors in high value resistance measurements in VNIIM

  • Author

    Semenov, Yu ; Samodurov, I.

  • Author_Institution
    VNIIM, St. Petersburg, Russia
  • fYear
    2010
  • fDate
    13-18 June 2010
  • Firstpage
    643
  • Lastpage
    644
  • Abstract
    VNIIM provides traceability of high-resistance measurements using single and transfer resistance standards comprising resistors wound with a cast microwire in glass tegument (CMW). Accuracy of 100 kOhm - 10 GOhm measurements is of ppm order and is confirmed by a number of key and bilateral comparisons. Resistance unit transfer from 1 GOhm to 1 (100) TOhm is carried out using T-shaped imitators made of microwire resistors.
  • Keywords
    electric resistance measurement; resistors; T-shaped imitators; VNIIM; cast microwire; glass tegument; microwire wound resistors; resistance 100 kohm to 10 Gohm; resistance measurements; resistance unit transfer; transfer resistance standards; Electrical resistance measurement; Electromagnetic measurements; Resistors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Precision Electromagnetic Measurements (CPEM), 2010 Conference on
  • Conference_Location
    Daejeon
  • Print_ISBN
    978-1-4244-6795-2
  • Type

    conf

  • DOI
    10.1109/CPEM.2010.5544296
  • Filename
    5544296