DocumentCode
3477898
Title
Cross-talk paths in array transducers
Author
Guess, J.F. ; Oakley, C.G. ; Douglas, S.J. ; Morgan, R.D.
Author_Institution
Sonovolt, Estes Park, CO, USA
Volume
2
fYear
1995
fDate
7-10 Nov 1995
Firstpage
1279
Abstract
Cross-talk is an important component of array transducer performance. It directly affects low level ring down, and limits aperture and steering angles. This paper describes a method of analysis which helps to sort out the sources and paths of acoustic cross-talk to allow improved array design. A linear array was built up in a step by step fashion with cross-talk measured at each step. Standard gain-phase testing and a novel cross-talk analysis method using time waveforms were both used to describe the cross-talk. Results demonstrate the contributions of cross-talk through the backing and through the kerf filler at each level of construction
Keywords
acoustic transducer arrays; biomedical equipment; biomedical ultrasonics; crosstalk; ultrasonic transducer arrays; analysis method; array transducer performance; cross-talk paths; kerf filler; medical diagnostic imaging; medical instrumentation; standard gain-phase testing; time waveforms; Acoustic pulses; Acoustic testing; Attenuation; Bandwidth; Ceramics; Couplings; Frequency response; Impedance; Ultrasonic transducer arrays; Vibrations;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 1995. Proceedings., 1995 IEEE
Conference_Location
Seattle, WA
ISSN
1051-0117
Print_ISBN
0-7803-2940-6
Type
conf
DOI
10.1109/ULTSYM.1995.495790
Filename
495790
Link To Document