DocumentCode :
3478400
Title :
A multi-step approach for accurate permittivity measurements of liquids using a transmission line method
Author :
Zhao, D. ; Rietveld, G. ; Teunisse, G.M.
Author_Institution :
Dutch Metrol. Inst., VSL, Delft, Netherlands
fYear :
2010
fDate :
13-18 June 2010
Firstpage :
767
Lastpage :
768
Abstract :
High-precision permittivity measurements of liquids are essential for specifying the specific absorption rate (SAR) of radio frequency power into human tissue. At VSL, a commercial unit is used for these measurements, but for reaching a low uncertainly some inherent disadvantages need to be overcome. In this paper, a multi-step approach is proposed using a combination of curve fitting, nonlinear progressive technique and an improved transmission line model. This approach is applicable even without a priori knowledge of the liquid properties. Using this approach, the applicable frequency range of the commercial unit is successfully extended from 200 MHz 3 GHz to at least 100 MHz-4 GHz, and the accuracy is improved from 2 % to 0.2 % 0.3 %.
Keywords :
biological techniques; curve fitting; permittivity measurement; transmission lines; SAR; VSL; curve fitting; frequency 100 MHz to 4 GHz; human tissue; liquids permittivity measurement; nonlinear progressive technique; radio frequency power; specific absorption rate; transmission line method; transmission line model; Dielectric liquids; Dielectric measurements; Electromagnetic measurements; Equations; Frequency measurement; Humans; Iterative algorithms; Permittivity measurement; Radio frequency; Slabs;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Precision Electromagnetic Measurements (CPEM), 2010 Conference on
Conference_Location :
Daejeon
Print_ISBN :
978-1-4244-6795-2
Type :
conf
DOI :
10.1109/CPEM.2010.5544368
Filename :
5544368
Link To Document :
بازگشت