DocumentCode :
3478590
Title :
A Neurostimulator Design for Long-term Animal Experiments
Author :
Zhou, Jingai ; Woo, Se Joon ; Park, Se Ik ; Lee, Seung Woo ; Chung, Hum ; Kim, Sung June
Author_Institution :
Sch. of Electr. Eng. & Comput. Sci., Seoul Nat. Univ., Seoul
fYear :
2007
fDate :
11-13 Oct. 2007
Firstpage :
453
Lastpage :
457
Abstract :
This article reports on a neural prosthesis stimulation system for long-term use in animal electrical stimulation experiments. The presented system consists of an implantable stimulator which provides continuous electrical stimulation, and an external component which provides preset stimulation parameters and power to the implanted stimulator via a paired RF (radio frequency) coil. A rechargeable internal battery and a parameter memory component were introduced to the implanted neural stimulator. As a result, the external component was not necessary during the stimulation cycles. The implantable stimulator was implemented with IC chips and the electronics, except for the stimulation electrodes, were hermetically packaged in a biocompatible metal case. A polyimide-based gold electrode array was used for realization of the animal implantation test using retinal prosthesis approach.
Keywords :
bioelectric potentials; biomedical electrodes; gold; neuromuscular stimulation; prosthetics; Au; IC chips; animal electrical stimulation; biocompatible metal case; implantable stimulator; neural prosthesis stimulation system; neurostimulator design; polyimide-based gold electrode array; retinal prosthesis; Animals; Batteries; Coils; Electrical stimulation; Electrodes; Electronics packaging; Hermetic seals; Integrated circuit packaging; Prosthetics; Radio frequency;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Frontiers in the Convergence of Bioscience and Information Technologies, 2007. FBIT 2007
Conference_Location :
Jeju City
Print_ISBN :
978-0-7695-2999-8
Type :
conf
DOI :
10.1109/FBIT.2007.67
Filename :
4524148
Link To Document :
بازگشت