Abstract
1.5D transducer arrays represent a promising improvement in ultrasonic B-mode imaging. The reduction of the slice thickness and the correction of phase aberration are potential applications of 1.5D arrays, while keeping the number of electronic channels on ultrasound beamformers between 64 and 256. The authors present the design and manufacturing, using a high density interconnection process, of a 1.5D curved linear array with 128×4 independent elements, allowing dynamic focusing in the elevation plane
Keywords
biomedical equipment; biomedical ultrasonics; ultrasonic transducer arrays; 128×4 channels 1.5D curved linear array; dynamic focusing; electronic channels; elevation plane; high density interconnection process; medical diagnostic imaging; phase aberration correction; slice thickness reduction; ultrasonic B-mode imaging; ultrasound beamformers; Acoustic scattering; Biomedical imaging; Focusing; Frequency; Gratings; Image sampling; Lenses; Performance analysis; Phased arrays; Sampling methods;