• DocumentCode
    3482715
  • Title

    Development of several sensors for Ca/sup ++/ potentiometric measurement in blood

  • Author

    Gouget, B. ; Gourmelin, Y. ; Feuillu, A. ; Truchaud, A.

  • fYear
    1988
  • fDate
    4-7 Nov. 1988
  • Firstpage
    734
  • Abstract
    An original protocol for the multicenter coordinated evaluation of seven analyzers for the measurement of ionized calcium is reported. The comparison of the instruments was made possible by using specific biological samples and newly developed HEPES or MOPS buffered aqueous reference solutions. The results for the evaluation of Ca/sup 2+/ instruments have shown that: (1) good precision and linearity are obtained for most of the analyzers with the aqueous buffered solutions; (2) serum is the most suitable medium; the values obtained on whole blood are more scattered and therefore whole blood specimens should be restricted for urgent circumstances; and (3) sampling and storage conditions of the specimens do not restrain routine determinations in clinical laboratories. This evaluation shows the good quality of small automatic instruments which allows rapid and reliable ionized calcium measurements.<>
  • Keywords
    biomedical measurement; blood; calcium; Ca/sup 2+/ potentiometric measurement; aqueous buffered solutions; blood; sensors; serum; small automatic instruments;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
  • Conference_Location
    New Orleans, LA, USA
  • Print_ISBN
    0-7803-0785-2
  • Type

    conf

  • DOI
    10.1109/IEMBS.1988.94983
  • Filename
    94983