DocumentCode :
3483037
Title :
Medical equipment sterilization using superhigh dose rate X-ray irradiation
Author :
Barinov, N.U. ; Dolgachev, G.I. ; Maslennikov, D.D. ; Nitishinsky, M.S. ; Ushakov, A.G. ; Zakatov, L.P. ; Zinoviev, O.A.
Author_Institution :
Res. Center, Kurchatov (I.V.) Inst. of Atomic Energy, Moscow, Russia
Volume :
2
fYear :
1998
fDate :
1998
Firstpage :
977
Abstract :
The results of sterilization tests made at the RS-20 generator of Bremsstrahlung for sterilizing purpose are reported. The measurements were done under different conditions in the area of 10-30 kGy dose levels at 1 MGy/s peak dose rate range for various microorganisms. The immediately irradiated volume was 0.036 of cubic meter with average dose rate at 1 kGy/hour provided in the center of the volume. The Bremsstrahlung was created by 2.0 MeV 10 kA 100 ns pulsed e-beam directed toward the target at 2 Hz. The tests were compared with those made simultaneously at continuous radionucleic radiation source using Co-60 isotope providing same dose values. It was found that full sterilization effect at RS-20 using 1 MGy/s peak dose rate takes 2-5 times lesser dose against radionucleic source and increases total efficiency of the system. Total dose decrease down to 5-12 kGy allows to reduce radiation loads on the equipment and broadens the area of sterilizing application
Keywords :
X-ray applications; biomedical equipment; bremsstrahlung; 10 kA; 100 ns; 2 Hz; 2 MeV; 5 to 30 kGy; Bremsstrahlung; RS-20 generator; average dose rate; medical equipment sterilization; microorganisms; pulsed e-beam; superhigh dose rate X-ray irradiation; Area measurement; Biomedical equipment; Decontamination; Electron beams; Influenza; Isotopes; Microorganisms; Physics; Power generation economics; Testing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
High-Power Particle Beams, 1998. BEAMS '98. Proceedings of the 12th International Conference on
Conference_Location :
Haifa
Print_ISBN :
0-7803-4287-9
Type :
conf
DOI :
10.1109/BEAMS.1998.817016
Filename :
817016
Link To Document :
بازگشت