DocumentCode :
348343
Title :
Automatic self-calibration of piezo systems in phase shifting Fizeau interferometers
Author :
Chu-Shik Kang ; Nicolaus, R.A. ; Bonsch, G.
Author_Institution :
Korea Res. Inst. of Stand. & Sci., Taejon, South Korea
Volume :
2
fYear :
1999
fDate :
Aug. 30 1999-Sept. 3 1999
Firstpage :
330
Abstract :
A digital piezo system comprises a PZT and a capacitive sensor combined in a single unit. The desired displacement can be achieved by simply inputting the displacement value into the software. However, when target value of /spl lambda//8 was input to the three PZT´s, the Fizeau plate did not show parallel movement, resulting in tilt to the linear fringes of the interferogram, In order to measure the length of the gauge block with high accuracy, this non-parallel movement of the Fizeau plate should be corrected precisely. This could be accomplished by multiplying correction factors to the target values of the piezo systems, which are self-measured by the phase shifting Fizeau interferometer.
Keywords :
calibration; light interferometers; piezoelectric actuators; PZT; automatic self-calibration; capacitive sensor; correction factors; digital piezo system; displacement value; gauge block; linear fringes; nonparallel movement; phase shifting Fizeau interferometers; software; target values; Calibration; Interference; Interferometers; Length measurement; Measurement standards; Optical interferometry; Performance evaluation; Phase shifters; Phase shifting interferometry; Volume measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Lasers and Electro-Optics, 1999. CLEO/Pacific Rim '99. The Pacific Rim Conference on
Conference_Location :
Seoul, South Korea
Print_ISBN :
0-7803-5661-6
Type :
conf
DOI :
10.1109/CLEOPR.1999.811438
Filename :
811438
Link To Document :
بازگشت