DocumentCode
3486025
Title
Phase aberration correction in two dimensions using a deformable array transducer
Author
Ries, Loriann L. ; Smith, Stephen W.
Author_Institution
Dept. of Biomed. Eng., Duke Univ., Durham, NC, USA
Volume
2
fYear
1995
fDate
7-10 Nov 1995
Firstpage
1439
Abstract
Two-dimensional arrays are required to implement an electronic two-dimensional phase correction in tissue. Instead of using a two-dimensional array, a new technique is proposed which uses a deformable transducer of significantly fewer channels for two-dimensional phase correction. Phase correction in azimuth is achieved by altering the electronic phase delay of the element. However, phase correction in elevation is achieved by tilting the element in elevation with a piezoelectric actuator. Here, simulations of the deformable array and results from a prototype deformable array are presented
Keywords
biomedical ultrasonics; medical image processing; ultrasonic transducer arrays; 2D phase aberration correction; azimuth; deformable array transducer; electronic two-dimensional phase correction; element tilting; medical diagnostic imaging; piezoelectric actuator; prototype deformable array; tissue; Acoustic beams; Azimuth; Biomedical engineering; Biomedical transducers; Delay; Error correction; Focusing; Phased arrays; Ultrasonic imaging; Ultrasonic transducer arrays;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 1995. Proceedings., 1995 IEEE
Conference_Location
Seattle, WA
ISSN
1051-0117
Print_ISBN
0-7803-2940-6
Type
conf
DOI
10.1109/ULTSYM.1995.495827
Filename
495827
Link To Document