Beat-to-beat prediction of left ventricular output during left ventricular bypass pumping
Author :
Sekii, Hiroyoshi ; Takano, Hisateru ; Taenaka, Yoshiyuki ; Takatani, Setsuo ; Noda, Hiroyuki ; Kinoshita, Masayuki ; Noda, Eisuke ; Yagura, Akihiko ; Akutsu, Tetsuzo
Author_Institution :
Nat. Cardiovascular Center, Osaka, Japan
fYear :
1988
fDate :
4-7 Nov. 1988
Firstpage :
1773
Abstract :
In order to estimate left-ventricular output from the pulse-pressure and pressure-flow relationship during left-ventricular-assist-device (LVAD) pumping, three methods were compared. The first method is based on the aortic compliance, while the second and third methods are based on the peripheral vascular resistance and the aortic impedance, respectively. Prior to clinical application, these methods were evaluated in a mock circulatory loop. The in vitro study revealed that in comparison to the electromagnetic flowmetering the second and third methods yielded higher correlation coefficients than the first method, indicating that they are more suitable for practical application during LVAD pumping.<>
Keywords :
cardiology; haemodynamics; patient treatment; aortic compliance; aortic impedance; cardiac haemodynamics; correlation coefficient; electromagnetic flowmetering; in vitro study; left ventricular bypass pumping; left ventricular output; mock circulatory loop; peripheral vascular resistance; pressure-flow relationship; pulse-pressure relationship;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
Conference_Location :
New Orleans, LA, USA
Print_ISBN :
0-7803-0785-2
Type :
conf
DOI :
10.1109/IEMBS.1988.94999
Filename :
94999
Link To Document :
بازگشت