DocumentCode :
3486428
Title :
Comparison of experimental techniques for evaluating the correction factor of a rectangular waveguide microcalorimeter
Author :
Cui, Xiaohai ; Crowley, T.P.
Author_Institution :
Nat. Inst. of Stand. & Technol., Boulder, CO, USA
fYear :
2010
fDate :
13-18 June 2010
Firstpage :
551
Lastpage :
552
Abstract :
We have evaluated the correction factor of a WR-15 (50 to 75 GHz) rectangular waveguide microcalorimeter with four different techniques. The four methods are in agreement. Our initial uncertainty analysis indicates that the technique with the lowest uncertainty is one that utilized two identical microcalorimeter core sections whose reference planes were connected to each other. This initial uncertainty is smaller than expected, which will require us to re-evaluate uncertainty contributions that were neglected.
Keywords :
calorimeters; measurement uncertainty; millimetre wave measurement; rectangular waveguides; WR-15 rectangular waveguide microcalorimeter; correction factor evaluation; frequency 50 GHz to 75 GHz; microcalorimeter core section; uncertainty analysis; Coaxial components; Electromagnetic heating; Electromagnetic measurements; Electromagnetic waveguides; Measurement standards; NIST; Performance evaluation; Power measurement; Radio frequency; Rectangular waveguides;