    Calibration of Rogowski coils at high currents

  • Author

    Draxler, K. ; Styblikova, R.

  • Author_Institution
    Fac. of Electr. Eng., Czech Tech. Univ. in Prague, Prague, Czech Republic
  • fYear
    2010
  • fDate
    13-18 June 2010
  • Firstpage
    537
  • Lastpage
    538
  • Abstract
    This paper describes the calibration of a Rogowski coil (RC) with an integrator assigned for current measurement in the range up to 25 kA at 50 Hz mains frequency. A current loop with 10 turns with a current of 2.5 kA is used for calibration. The magnetic field generated in such a way differs from the field of one turn placed in the centre and passed with a current of 25 kA. The influence of non-homogeneities of winding and the cross-section of the RC may cause a calibration error when using a current loop. The difference in the results between using a current loop and one turn and the measurement uncertainties are given below.
  • Keywords
    calibration; coils; measurement uncertainty; RC cross section; RC winding nonhomogeneity; Rogowski coil calibration; calibration error; current 2.5 kA; current 25 kA; current loop; current measurement; frequency 50 Hz; integrator; magnetic field; measurement uncertainty; Calibration; Coils;