DocumentCode :
3488418
Title :
Design of an interstitial capacitive hyperthermia system operating at 27.12 MHz
Author :
Nadi, M. ; Marchal, C. ; Prieur, G. ; Tosser, A. ; Mabire, J.P.
Author_Institution :
Univ. of Sci., Nancy, France
fYear :
1988
fDate :
4-7 Nov. 1988
Firstpage :
1275
Abstract :
The authors suggest an alternative interstitial method operating at 27.12 MHz. Physically, the system is essentially capacitive and the heating is induced by the conduction current between the ´hot´ electrode and the ground plane. Thermographic results obtained with a two-channel homemade apparatus are presented. The simplicity and low cost of the technology, the possibility of tailoring the electrodes to the geometry and the tumor site, and the radial and longitudinal uniformity of the heating represent the major advantages of this method. A commercial prototype with eight electrodes is presented.<>
Keywords :
biomedical equipment; biothermics; radiation therapy; 2-channel homemade apparatus; 27.12 MHz; conduction current; electrodes; ground plane; heating uniformity; hot electrode; interstitial capacitive hyperthermia system; medical instrumentation; thermography;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
Conference_Location :
New Orleans, LA, USA
Print_ISBN :
0-7803-0785-2
Type :
conf
DOI :
10.1109/IEMBS.1988.95015
Filename :
95015
Link To Document :
بازگشت