Medical device EMI: FDA analysis of incident reports, and recent concerns for security systems and wireless medical telemetry
Author
Witters, Donald ; Portnoy, S. ; Casamento, Jon ; Ruggera, Paul ; Bassen, Howard
Author_Institution
Center for Devices & Radiological Health, US Food & Drug Adm., Rockville, MD, USA
Volume
2
fYear
2001
fDate
2001
Firstpage
1289
Abstract
FDA has evaluated reports of medical device malfunctions caused by electromagnetic interference (EMI), performed device testing, and developed standardized test procedures. Over 500 incident reports are suspected to be attributable to EMI affecting cardiac devices. More than 80 of these reports involve cardiac and other medical device interactions with electronic security systems. EMI presents a risk to patient safety and medical device effectiveness that is likely to continue as the use of electromagnetic energy in the medical device environment increases (e.g., cell phones, security systems). Developments can reduce these risks, such as the allocation of dedicated frequency bands for the new wireless medical telemetry service (WMTS) designed to protect transmissions of patient vital signs from interference by other intentional transmitters
Keywords
biomedical equipment; biomedical telemetry; electromagnetic compatibility; electromagnetic interference; security; EMC; FDA; cardiac devices; cell phones; dedicated frequency bands; electromagnetic compatibility; electromagnetic energy; electromagnetic interference; incident reports analysis; medical device EMI; medical device effectiveness; medical device malfunctions; medical device testing; patient safety; patient vital signs transmission; security systems; standardized test procedures; wireless medical telemetry; Cellular phones; Communication system security; Electromagnetic devices; Electromagnetic interference; Medical tests; Performance evaluation; Protection; Radio spectrum management; Safety devices; Telemetry;
fLanguage
English
Publisher
ieee
Conference_Titel
Electromagnetic Compatibility, 2001. EMC. 2001 IEEE International Symposium on
Conference_Location
Montreal, Que.
Print_ISBN
0-7803-6569-0
Type
conf
DOI
10.1109/ISEMC.2001.950633
Filename
950633
Link To Document