DocumentCode
3497154
Title
On the evaluation a an ultrasonic transducer response using a fixing angle device in the acquisition of Doppler blood flow signals during coronary graft surgery
Author
Gonzalez, Javier Serrano ; Dominguez, I.S. ; Hernandez, M.V. ; Cruz, M.F. ; Nocetti, D.F.G. ; Martínez, Juan Prohías ; Inclán, Alejandro Villar ; Salas, Lorenzo Leija
Author_Institution
Inst. de Investig. en Mat. Aplic. y en Sist., Univ. Nac. Autonoma de Mexico, Mexico City, Mexico
fYear
2011
fDate
March 28 2011-April 1 2011
Firstpage
148
Lastpage
151
Abstract
This paper proposes the design and use of a device which will fix the angle of an ultrasonic transducer used in the acquisition of Ultrasonic Doppler signals during coronary graft surgery. The aim of this device is to act as a tool which will guide the surgeon to position the ultrasonic transducer in the correct angle and plane which is essential on the calculation of the blood flow through the implanted graft. An experimental set-up has been implemented in order to evaluate the transducer response when used with the device. The radiation fields and attenuation due to the presence of the device are discussed and analyzed. Results show the effectiveness of the device in handling the transducer and warranty the quality of the acquired signals and measurements.
Keywords
biomedical transducers; biomedical ultrasonics; blood flow measurement; cardiovascular system; surgery; ultrasonic transducers; Doppler blood flow signals; coronary graft surgery; fixing angle device; implanted graft; radiation fields; ultrasonic transducer; Acoustics; Blood flow; Conferences; Couplings; Hospitals; IEEE catalog; Blood-flow Measurement; Coronary Graft Surgery; Doppler Effect; Ultrasound Transducers;
fLanguage
English
Publisher
ieee
Conference_Titel
Health Care Exchanges (PAHCE), 2011 Pan American
Conference_Location
Rio de Janeiro
Print_ISBN
978-1-61284-915-7
Type
conf
DOI
10.1109/PAHCE.2011.5871868
Filename
5871868
Link To Document