• DocumentCode
    3499498
  • Title

    Impedance biosensor based on interdigitated electrode arrays for detection of low levels of E.coli O157:H7

  • Author

    Dastider, S.G. ; Barizuddin, S. ; Wu, Yaowu ; Dweik, M. ; Almasri, M.

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Univ. of Missouri, Columbia, MO, USA
  • fYear
    2013
  • fDate
    20-24 Jan. 2013
  • Firstpage
    955
  • Lastpage
    958
  • Abstract
    We report an interdigitated electrode array (IDEA) impedance biosensor with rapid detection of Escherichia coli O157:H7. This biosensor has significantly better detection capability compared with the conventional microbiological testing method. The total detection time for this biosensor is under 3 hours. Dielectrophoresis (DEP) is used in conjunction with impedance spectroscopy to improve the detection capability of the biosensor. The IDEA surface is functionalized using anti-E.coli antibody for selectivity of the bacteria. The dose response was in-between 3 × 106 CFU/mL and 3 × 102 CFU/mL. This is 10 fold better than the detection limit previously reported using impedance based biosensors.
  • Keywords
    biosensors; electrochemical electrodes; electrochemical impedance spectroscopy; electrophoresis; microorganisms; Escherichia coli O157:H7 detection; antiE.coli antibody; conventional microbiological testing method; detection capability; dielectrophoresis; dose response; impedance biosensor; impedance spectroscopy; interdigitated electrode arrays; time 3 h; total detection time; Biosensors; Electrodes; Focusing; Impedance; Microchannel; Microorganisms; Surface impedance;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Micro Electro Mechanical Systems (MEMS), 2013 IEEE 26th International Conference on
  • Conference_Location
    Taipei
  • ISSN
    1084-6999
  • Print_ISBN
    978-1-4673-5654-1
  • Type

    conf

  • DOI
    10.1109/MEMSYS.2013.6474404
  • Filename