DocumentCode :
35016
Title :
A Pneumatically Actuated Target Stabilization Device for MRI-Guided Breast Biopsy
Author :
Iranpanah, Behzad ; Chen, Maggie ; Patriciu, Alexandru ; Sirouspour, Shahin
Author_Institution :
Dept. of Electr. & Comput. Eng., McMaster Univ., Hamilton, ON, Canada
Volume :
20
Issue :
3
fYear :
2015
fDate :
Jun-15
Firstpage :
1288
Lastpage :
1300
Abstract :
This paper introduces a novel tissue stabilization device for magnetic resonance imaging (MRI)-guided breast biopsy. The device is comprised of two pneumatically actuated support plates that would stabilize the biopsy target movements during needle insertion. An optimized geometry for the support plates allows for a good degree of tissue stabilization without relying on large compression forces. The plate configuration can also be adjusted inside the magnet bore using pneumatic actuators driven by pressure-controlled valves that are placed in the MRI control room. This capability allows for the compensation of the target displacement based on MRI image feedback. When combined with a separate needle drive mechanism, this stabilization device would enable in-bore MRI-guided breast biopsy. Experiments performed on chicken breast tissue with a prototype of the device demonstrate the effectiveness of this mechanism in increasing needle targeting accuracy using two simple error correction strategies. Furthermore, MRI compatibility tests are carried out to assess the performance of the device inside MRI.
Keywords :
biological tissues; biomedical MRI; medical control systems; medical image processing; needles; pneumatic actuators; pressure control; stability; MRI compatibility tests; MRI control room; MRI image feedback; biopsy target movement stabilization; chicken breast tissue; error correction strategies; in-bore MRI-guided breast biopsy; magnet bore; magnetic resonance imaging-guided breast biopsy; needle drive mechanism; needle insertion; performance assessment; plate configuration; pneumatic actuators; pneumatically actuated support plates; pneumatically actuated target stabilization device; pressure-controlled valves; tissue stabilization device; Biopsy; Breast biopsy; Geometry; Magnetic resonance imaging; Needles; Shape; Breast biopsy; MRI compatibility; MRI-guided biopsy; compression plates; magnetic resonance imaging (MRI); pneumatic actuators; tissue stabilization;
fLanguage :
English
Journal_Title :
Mechatronics, IEEE/ASME Transactions on
Publisher :
ieee
ISSN :
1083-4435
Type :
jour
DOI :
10.1109/TMECH.2014.2341657
Filename :
6880335
Link To Document :