DocumentCode
350296
Title
A fast image registration technique for motion artifact reduction in DSA
Author
Meijering, Erik H W ; Zuiderveld, Karel J. ; Niessen, Wiro J. ; Viergever, Max A.
Author_Institution
Image Sci. Inst., Utrecht Univ., Netherlands
Volume
3
fYear
1999
fDate
1999
Firstpage
435
Abstract
In digital subtraction angiography (DSA), patient motion is the primary cause of image quality degradation. The motion correction algorithms developed so far were not sufficiently fast so as to be suitable for integration in a clinical setting. In this paper we describe a new image registration technique for motion artifact reduction in DSA which is fully automatic, effective, and computationally very efficient. Using an image content driven control point selection mechanism and modern graphics hardware for image warping, the algorithm requires less than one second per DSA image (on average). Preliminary experiments on cerebral DSA images illustrate the applicability of the technique
Keywords
biomedical MRI; image registration; image sequences; motion estimation; cerebral DSA images; clinical setting; digital subtraction angiography; graphics hardware; image content driven control point selection mechanism; image quality degradation; image registration technique; image warping; motion artifact reduction; motion correction algorithms; patient motion; Angiography; Automatic control; Degradation; Graphics; Hardware; Image quality; Image registration; Uniform resource locators; Visualization; X-ray imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Image Processing, 1999. ICIP 99. Proceedings. 1999 International Conference on
Conference_Location
Kobe
Print_ISBN
0-7803-5467-2
Type
conf
DOI
10.1109/ICIP.1999.817151
Filename
817151