• DocumentCode
    3504742
  • Title

    Computer simulation of Purkinje-Sanson images for the biofeedback therapy of myopia

  • Author

    Fukumoto, Ichiro

  • Author_Institution
    Inst. of Biomed. Eng., Nagaoka Univ. of Technol., Niigata, Japan
  • Volume
    2
  • fYear
    1999
  • fDate
    36495
  • Firstpage
    1166
  • Abstract
    In order to treat pseudo-myopia, the authors have proposed a biofeedback therapy system with real time feedback of a patient´s eye-lens thickness using Purkinje-Sanson (PS) images. The PS images are recorded with an infrared video-camera during lightening by visible LED light. After the images are digitized by a personal computer, the PS parameters (distances between the images or the area of the images) are calculated in order to estimate the lens thickness. The measured data show good coincidence with the anatomical data of extracted pigs eyes. Several clinical measurements on normal and nearsighted subjects show good correlations with r=0.958 between the real distances from the objects and the PS parameters. Computer simulations based on the optical model of the PS images are executed too, which show relative good coincidence with the real object´s distances
  • Keywords
    digital simulation; eye; feedback; infrared imaging; medical image processing; patient treatment; thickness measurement; vision defects; Purkinje-Sanson images; biofeedback therapy; computer simulation; digitized images; infrared video-camera; myopia; patient´s eye-lens thickness; visible LED light; Biological control systems; Computer simulation; Data mining; Feedback; Infrared imaging; Lenses; Light emitting diodes; Medical treatment; Microcomputers; Real time systems;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    TENCON 99. Proceedings of the IEEE Region 10 Conference
  • Conference_Location
    Cheju Island
  • Print_ISBN
    0-7803-5739-6
  • Type

    conf

  • DOI
    10.1109/TENCON.1999.818633
  • Filename
    818633