DocumentCode :
3524058
Title :
A dual-user teleoperated system with Virtual Fixtures for robotic surgical training
Author :
Shahbazi, Moein ; Atashzar, S. Farokh ; Patel, Rajni V.
Author_Institution :
Canadian Surg. Technol. & Adv. Robot. (CSTAR), Lawson Health Res. Inst., London, ON, Canada
fYear :
2013
fDate :
6-10 May 2013
Firstpage :
3639
Lastpage :
3644
Abstract :
This paper proposes a teleoperated dual-user system incorporating Virtual Fixtures (VFs) that allows concurrent performance of a robotic surgical task by an expert and a trainee. In order to guide the trainee through the procedure, an adaptive VF is created in the trainee´s workspace according to the motion generated by the expert who is performing the surgery at the same time. The VF gets adaptively adjusted based on the level of expertise the trainee shows during the surgery. In addition, the trainee´s level of expertise is used to adaptively adjust the dual-user dominance factor in an online fashion, which gives the trainee some authority over the task based on his/her skill level. To quantify the trainee´s expertise level, a performance measure is proposed, based on the force generated by the VF. Three performance measures from the literature are also used. To satisfy the desired objectives of the proposed system, an impedance-based control methodology is adopted. Stability of the closed-loop system is investigated using the small-gain theorem. A sufficient stability condition is derived that guarantees stability in the presence of time-varying communication delay. Experimental results are given to validate the performance of the system.
Keywords :
closed loop systems; electric variables control; medical robotics; stability; surgery; telerobotics; adaptive VF; closed-loop system; dual-user dominance factor; dual-user teleoperated system; impedance-based control methodology; robotic surgical task; robotic surgical training; small-gain theorem; stability condition; time-varying communication delay; trainee workspace; virtual fixtures; Fixtures; Force; Measurement; Robots; Stability analysis; Surgery; Training; Dual-User System; Robotic Surgical Training; Teleoperation; Virtual Fixture;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Robotics and Automation (ICRA), 2013 IEEE International Conference on
Conference_Location :
Karlsruhe
ISSN :
1050-4729
Print_ISBN :
978-1-4673-5641-1
Type :
conf
DOI :
10.1109/ICRA.2013.6631088
Filename :
6631088
Link To Document :
بازگشت