• DocumentCode
    3525854
  • Title

    On-line system for automated auditory evoked response threshold determination

  • Author

    Delgado, R.E. ; Özdamar, Ö ; Miskiel, E.

  • Author_Institution
    Dept. of Biomed. Eng., Miami Univ., Coral Gables, FL, USA
  • fYear
    1988
  • fDate
    4-7 Nov. 1988
  • Firstpage
    1472
  • Abstract
    An online computer system has been developed to determine auditory-evoked-response thresholds for objective hearing testing of otherwise untestable patients. The system uses click-evoked auditory brainstem responses. Online threshold determination is accomplished by two modules: a response recognition unit which determines if a response is present and a threshold tracking unit which utilizes a modified PEST (parameter estimation by sequential testing) algorithm to determine the click intensity testing level. Thresholds obtained manually and automatically are compared in 161 tests from 62 ears. Results indicate that automatically obtained thresholds show a strong agreement with those obtained visually as determined by a panel of experts.<>
  • Keywords
    bioelectric potentials; biomedical measurement; hearing; medical diagnostic computing; automated auditory evoked response threshold determination; click-evoked auditory brainstem responses; objective hearing testing; online computer system; parameter estimation by sequential testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 1988. Proceedings of the Annual International Conference of the IEEE
  • Conference_Location
    New Orleans, LA, USA
  • Print_ISBN
    0-7803-0785-2
  • Type

    conf

  • DOI
    10.1109/IEMBS.1988.95333
  • Filename
    95333