• DocumentCode
    3526409
  • Title

    Beam profile monitoring system for proton therapy

  • Author

    Ho, Chili ; Chen, Augustine E. ; Dai, Shujhen ; Chu, Minglee ; Lin, Chih-Hsun ; Teng, Ping-Kun ; Wang, Chung-Hsiang ; Chao, Tsi-Chian ; Lee, Chung-Chi ; Tung, Chuang-Jong ; Duh, Ting-Shien

  • Author_Institution
    Dept. of Phys., Nat. Central Univ., Jhongli, Taiwan
  • fYear
    2010
  • fDate
    Oct. 30 2010-Nov. 6 2010
  • Firstpage
    440
  • Lastpage
    442
  • Abstract
    An innovative design of 16-channel front-end electronics for ionization chamber in monitoring proton beam profile has been developed to feature simplicity, low cost, durability, and reliability without compromising its performances. All the parts are available off the shelf. The front-end electronics board is powered by USB connection, through which the data are transferred to the host computer for storage and further online analysis. The system has been studied with Geant4 MC and tested with both electron and proton beams.
  • Keywords
    Monte Carlo methods; biomedical electronics; electron beams; ionisation chambers; proton beams; radiation therapy; Geant4 MC; USB connection; electron beams; front-end electronics; ionization chamber; proton beam profile monitoring; proton beams; proton therapy; Argon; Current measurement; Integrated circuits; Ionization chambers; Monitoring; Particle beams; Protons;