Title :
Application of Monte Carlo simulation to 111In standardization by means of a 4πβ-γ coincidence system
Author :
Koskinas, M.F. ; Brito, A.B. ; Takeda, M.N. ; Dias, M.S.
Author_Institution :
Nucl. Metrol. Lab., IPEN - Inst. de Pesquisas Energeticas e Nucl., São Paulo, Brazil
fDate :
Oct. 30 2010-Nov. 6 2010
Abstract :
In this paper the application of Monte Carlo simulation for standardizing 111In in a 4π(PC)-NaI(Tl) coincidence system is presented. The Monte Carlo simulation is applied using ESQUEMA code, which was developed by the Nuclear Metrology Laboratory (LMN) in IPEN, Sao Paulo, Brazil. The experimental results obtained by means a linear fitting were compared with those determined by the Monte Carlo prediction and they are in good agreement within the uncertainty.
Keywords :
Monte Carlo methods; radioisotopes; scintillation counters; standardisation; 4pion-NaI(Ti) coincidence system; 111ln standardization; ESQUEMA code; Monte Carlo prediction; Monte Carlo simulation; Extrapolation; Fitting; Laboratories; Metrology; Monte Carlo methods; Radiation detectors; Uncertainty;
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
Conference_Location :
Knoxville, TN
Print_ISBN :
978-1-4244-9106-3
DOI :
10.1109/NSSMIC.2010.5873806