Abstract :
Transcutaneous energy transmission systems for the artificial heart with synchronous rectifiers were developed. The input voltage is 24 V and the output voltage is 9 to 18 V. This paper shows the efficiency of the system with synchronous rectifier using a digital PLL technique achieved 92% DC-DC efficiency at 5 mm coil distance. Using the ferrite core and a digital PLL technique, inductor and capacitor were eliminated from the rectifier circuit. And the circuit was mounted to the back of the secondary coil. And we developed the control systems for stable output voltage and apply this result to other contact power supply system.
Keywords :
capacitors; digital phase locked loops; inductors; power supplies to apparatus; prosthetic power supplies; rectification; rectifying circuits; 24 V; 5 mm; 9 to 18 V; 92 percent; artificial heart; capacitor; coil; contactless power supplies; digital PLL technique; ferrite core; inductor; synchronous rectification; synchronous rectifiers; transcutaneous energy transmission systems; Artificial heart; Capacitors; Circuits; Coils; Ferrites; Inductors; Phase locked loops; Power supplies; Rectifiers; Voltage;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Magnetics Conference, 2005. INTERMAG Asia 2005. Digests of the IEEE International
Print_ISBN :
0-7803-9009-1
Type :
conf
DOI :
10.1109/INTMAG.2005.1463543
Filename :
1463543
Link To Document :
بازگشت