Title :
Characterization of low energy Lu background on continuous LYSO blocks
Author :
Lois, Cristina ; Aguiar, Pablo ; Couce, Beatriz ; Iglesias, Alfredo
Author_Institution :
Facultade de Fis., Univ. de Santiago de Compostela, Santiago de Compostela, Spain
fDate :
Oct. 30 2010-Nov. 6 2010
Abstract :
The presence of the naturally occurring isotope 176Lu gives rise to background count rates in LSO/LYSO crystals to be used in PET and SPECT systems. The aim of this work is to measure 176Lu background count rates and its spatial distributions in a continuous LYSO block. Our results show that the low energy component of 176Lu background is related to partial energy deposition mainly produced near the crystal edge showing a hot-perimeter artifact. Furthermore, 176Lu background rate is observed to be ×10 times lower for a 100-180 keV than for a 411-611 keV energy window. Our findings indicate that when using monolithic LYSO crystal blocks, 176Lu background is not expected to significantly deteriorate the performance of single photon detection systems.
Keywords :
crystals; isotopes; lutetium; positron emission tomography; single photon emission computed tomography; Lu; PET; SPECT; background count rates; continuous LYSO blocks; crystals; low energy Lu background; low energy component; naturally occurring isotope; partial energy deposition; single photon detection systems; spatial distributions; Cameras; Crystals; Detectors; Photonics; Positron emission tomography; Single photon emission computed tomography; 176Lu intrinsic radioactivity; PET/SPECT detectors; instrumentation;
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
Conference_Location :
Knoxville, TN
Print_ISBN :
978-1-4244-9106-3
DOI :
10.1109/NSSMIC.2010.6036244