DocumentCode :
3532261
Title :
A common approach to image reconstruction for different applications of Compton cameras
Author :
Schöne, Sebastian ; Shakirin, Georgy ; Kormoll, Thomas ; Herbach, Claus-Michael ; Pausch, Guntram ; Enghardt, Wolfgang
Author_Institution :
Dept. of Radiat. Phys., Res. Center Dresden-Rossendorf, Dresden, Germany
fYear :
2010
fDate :
Oct. 30 2010-Nov. 6 2010
Firstpage :
2292
Lastpage :
2293
Abstract :
A unified method for representation and calculation of the system matrix for Compton cameras has been proposed. The method defines first a photon interaction scheme in form of a logic photon state machine and then calculates the system matrix elements based on the transition probabilities for each state. Additionally uncertainties of the measurement can be taken into account in form of probability density distributions or, in case of simplifications, as scalars. The method provides a structured approach for the system matrix calculation which simplifies the software development process and the optimization of the computational performance. Additionally, the system matrix calculation is completely independent of the reconstruction algorithm itself. Therefore, the same code can be used for the reconstruction of the measured data collected with the different hardware and for different applications (i.e. medical imaging or homeland security).
Keywords :
finite state machines; image reconstruction; medical image processing; optimisation; probability; radioisotope imaging; Compton cameras; image reconstruction; logic photon state machine; optimization; photon interaction scheme; probability density distribution; software development process; system matrix; transition probabilities;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
Conference_Location :
Knoxville, TN
ISSN :
1095-7863
Print_ISBN :
978-1-4244-9106-3
Type :
conf
DOI :
10.1109/NSSMIC.2010.5874192
Filename :
5874192
Link To Document :