Title :
Hardware setup for the next generation of 3D Ultrasound Computer Tomography
Author :
Gemmeke, H. ; Berger, L. ; Birk, M. ; Göbel, G. ; Menshikov, A. ; Tcherniakhovski, D. ; Zapf, M. ; Ruiter, N.V.
Author_Institution :
Inst. for Data Process. & Electron., Karlsruhe Inst. of Technol., Eggenstein-Leopoldshafen, Germany
fDate :
Oct. 30 2010-Nov. 6 2010
Abstract :
We describe the second generation of a 3D-Ultrasound Computer Tomography (USCT) system. After we achieved in the first generation a device with sub-wavelength resolution and three imaging modalities (reflection, attenuation, speed of sound) and tested it with static phantoms, we developed a device for in-vivo imaging. In the new system the geometry of transducers and their spatial distribution is optimized in respect to uniformity and high value of: contrast, resolution, and illumination. Furthermore we developed new electronics which allows faster DAQ (≤ 2 min) and contains larger and faster FPGAs to use their processing power for data pre-processing.
Keywords :
biomedical transducers; biomedical ultrasonics; computerised tomography; field programmable gate arrays; phantoms; 3D ultrasound computer tomography; FPGA; data pre-processing; hardware setup; in-vivo imaging; static phantoms; sub-wavelength resolution; transducers; Apertures; Data acquisition; Field programmable gate arrays; Imaging; Receivers; Three dimensional displays; Transducers; Computer tomography; Medical imaging equipment; Ultrasonography;
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
Conference_Location :
Knoxville, TN
Print_ISBN :
978-1-4244-9106-3
DOI :
10.1109/NSSMIC.2010.5874228