Title :
High resolution image reconstruction with constrained, total-variation minimization
Author :
Sidky, Emil Y. ; Chartrand, Rick ; Duchin, Yuval ; Ullberg, Christer ; Pan, Xiaochuan
Author_Institution :
Dept. of Radiol., Univ. of Chicago, Chicago, IL, USA
fDate :
Oct. 30 2010-Nov. 6 2010
Abstract :
This work is concerned with applying iterative image reconstruction, based on constrained total-variation minimization, to low-intensity X-ray CT systems that have a high sampling rate. Such systems pose a challenge for iterative image reconstruction, because a very fine image grid is needed to realize the resolution inherent in such scanners. These image arrays lead to under-determined imaging models whose inversion is unstable and can result in undesirable artifacts and noise patterns. There are many possibilities to stabilize the imaging model, and this work proposes a method which may have an advantage in terms of algorithm efficiency. The proposed method introduces additional constraints in the optimization problem; these constraints set to zero high spatial frequency components which are beyond the sensing capability of the detector. The method is demonstrated with an actual CT data set and compared with another method based on projection up-sampling.
Keywords :
computerised tomography; image reconstruction; image sampling; iterative methods; medical image processing; minimisation; constrained total variation minimization; high resolution image reconstruction; image grid; iterative image reconstruction; low-intensity X-ray CT systems; optimization; projection up-sampling; Computed tomography; Detectors; Image reconstruction; Image resolution; Pixel; Variable speed drives;
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
Conference_Location :
Knoxville, TN
Print_ISBN :
978-1-4244-9106-3
DOI :
10.1109/NSSMIC.2010.5874263