    New calibration and evaluation method for PET scanners using point-like radioactive sources

  • Author

    Hasegawa, Tomoyuki ; Oda, Keiichi ; Sato, Yasushi ; Wada, Yasuhiro ; Yamada, Takahiro ; Yoshida, Eiji ; Murayama, Hideo ; Saito, Kyoko ; Takeda, Toru ; Kikuchi, Kei

  • Author_Institution
    Sch. of Allied Health Sci., Kitasato Univ., Sagamihara, Japan
  • fYear
    2010
  • fDate
    Oct. 30 2010-Nov. 6 2010
  • Firstpage
    2929
  • Lastpage
    2930
  • Abstract
    Objective: In order to improve the convenience and reliability of the calibration procedure with respect to the quantitative aspects of PET scanners, we have been developing new methods that rely on the use of point-like radioactive sources. The aim of this study is to report the results of the application of the proposed method for obtaining the cross-calibration factors (CCFs) of a clinical PET scanner for the first time. Methods: A 22Na point-like source was moved in the axial direction to cover the axial field-of-view (FOV) of a commercial PET scanner, SET-2400W (Shimadzu). From reconstructed images, saturated regions-of-interest (ROI) values were used to calculate the CCFs that were defined slice by slice. Results: It was possible to obtain CCFs that were practically equivalent to those obtained by the standard method with cylindrical water phantoms. The variations in the CCFs obtained over about a half-year were well reproduced by the proposed method. Conclusion: The proposed calibration method, which is based on the use of a point-like radioactive source, is useful for determining CCFs and their variations for a clinical PET scanner.
  • Keywords
    calibration; image reconstruction; medical image processing; phantoms; positron emission tomography; PET scanners; SET-2400W; axial field-of-view; calibration; cross-calibration factors; cylindrical water phantoms; image reconstruction; point-like radioactive sources; regions-of-interest values; Calibration; Detectors; Phantoms; Positron emission tomography; Standardization; Three dimensional displays;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
  • Conference_Location
    Knoxville, TN
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-9106-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2010.5874332
  • Filename
    5874332