Experimental feasibility of multi-material decomposition imaging in small animal SPECT/CT system
Author
Cho, Hyo-Min ; Kim, Hee-Joung ; Lee, Chang-Lae ; Pivovaroff, Michael ; Seo, Youngho
Author_Institution
Dept. of Radiol. & Biomed. Imaging, Univ. of California, San Francisco, CA, USA
fYear
2010
fDate
Oct. 30 2010-Nov. 6 2010
Firstpage
3187
Lastpage
3189
Abstract
The photon counting detectors such as cadmium zinc telluride (CZT) and cadmium telluride (CdTe) have powerful advantages compared to energy integrating detectors. CZT or CdTe can detect individual gamma-ray or x-ray photon with energy discrimination. In recent years, energy-resolving and material decomposition x-ray imaging based on photon counting detectors has attracted attention from biomedical imaging researchers. We evaluated a large-area (20 cm × 20 cm) CZT detector originally built as a radionuclide detector for a small animal SPECT/CT system in combination of a microfocus x-ray source with a general goal of developing a material-decomposition imaging method. In this paper, we present experimental results from a feasibility study of multi-material decomposition imaging using the developed small animal SPECT/CT system. Our small animal SPECT/CT system has the unique capability to arrange detectors and sources flexibly. For the multi-material decomposition scan, the CZT detector and the x-ray tube were re-arranged in line. List-mode data of a phantom containing 7 different materials, with a range of densities and atomic numbers, illuminated by x-ray were acquired. The x-ray exposure conditions were: 50 kVp and 0.5 mA with 1.5-mm Al filtration. The acquired image was corrected for bad and hot pixels, gain, and offset. The attenuation profiles of each material were calculated against the x-ray energy. Measured attenuation profiles of each material were in a good agreement with the reference data of the NIST physics laboratory. In this study, we demonstrated that multi-material decomposition imaging is experimentally feasible using the photon-counting CZT detector and polychromatic x-ray. Since our system allows rotation and a large active area of CZT, we will acquire material-decomposition data tomographically in the near future.
Keywords
X-ray detection; cadmium compounds; gamma-ray detection; phantoms; photon counting; semiconductor counters; single photon emission computed tomography; solid scintillation detectors; CZT detector; CdTe; CdZn; X-ray photon; biomedical imaging; cadmium telluride; cadmium zinc telluride; gamma-ray; multimaterial decomposition imaging; phantom; photon counting detectors; radionuclide detector; small animal SPECT/CT system; Attenuation; Computed tomography; Detectors; Materials; Photonics; X-ray imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
Conference_Location
Knoxville, TN
ISSN
1095-7863
Print_ISBN
978-1-4244-9106-3
Type
conf
DOI
10.1109/NSSMIC.2010.5874392
Filename
5874392
Link To Document