Title :
Co-fan-sum ratio algorithm for randoms smoothing and detector normalization in PET
Author :
Watson, Charles C.
Author_Institution :
Siemens Healthcare Mol. Imaging, Knoxville, TN, USA
fDate :
Oct. 30 2010-Nov. 6 2010
Abstract :
We describe a new exact analytical algorithm for estimating the single event rates in detectors from random coincidence data (for randoms variance reduction), and detector efficiencies from true coincidence data (for normalization), in positron emission tomography (PET). The estimates are derived from the ratios of the co-fan sums of coincidence events between individual detectors. The co-fan for any two detectors is defined as the set of co-detectors they have in common among the measured lines of response. The estimates are unbiased and have noise properties similar to fan-sum estimates. The detector efficiency estimation algorithm additionally employs the concept of a mask-restricted co-fan sum that can be easily adapted to a known source distribution, and thus does not require uniform illumination of all lines of response, or a centered uniform cylindrical source, for accurate estimation. The algorithm is simple to implement and can be applied in 2D or 3D. Some examples are given.
Keywords :
data acquisition; medical image processing; phantoms; positron emission tomography; random processes; smoothing methods; PET; co-fan-sum ratio algorithm; codetectors; data acquisition; detector efficiency estimation algorithm; detector normalization; phantom; random smoothing; Approximation algorithms; Detectors; Estimation; Noise; Phantoms; Positron emission tomography; Three dimensional displays;
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
Conference_Location :
Knoxville, TN
Print_ISBN :
978-1-4244-9106-3
DOI :
10.1109/NSSMIC.2010.5874420