Towards a high count rate energy resolving CdTe Hybrid pixel detector
Author
Berry, A. ; Lynch, A. ; Nikulin, D. ; Ruat, M. ; Fitrio, D. ; Tjoa, S. ; King, S. ; Mujcinovic, E. ; Mohan, A. ; Lewis, R.
Author_Institution
Sch. of Phys., Monash Univ., Clayton, VIC, Australia
fYear
2010
fDate
Oct. 30 2010-Nov. 6 2010
Firstpage
3825
Lastpage
3829
Abstract
A prototype CdTe pixel detector system with 19 pixels arranged in a 200um pitch hexagonal format is being developed for energy resolved medical X-ray imaging applications; this device should be capable of detecting input photon rates of up to 2*108 counts per mm2 and energy resolution of approximately 5% at 122keV. This paper presents a description of this system along with the initial spectral and count rate results. A comparison is made with a similar single 2*2mmpixel 700um thick CdTe detector.
Keywords
X-ray imaging; computerised tomography; semiconductor counters; CdTe hybrid pixel detector; count rate results; high count rate energy; medical X-ray imaging applications; photon rates; pitch hexagonal format; Arrays; Detectors; Energy resolution; Materials; Photonics; Pixel; Prototypes;