• DocumentCode
    3536703
  • Title

    Photon-counting energy-resolving CdTe detectors for high-flux x-ray imaging

  • Author

    Barber, William C. ; Nygard, Einar ; Wessel, Jan C. ; Malakhov, Nail ; Hartsough, Neal E. ; Gandhi, Thulasi ; Wawrzyniak, Gregor ; Iwanczyk, Jan S.

  • Author_Institution
    DxRay, Inc., Northridge, CA, USA
  • fYear
    2010
  • fDate
    Oct. 30 2010-Nov. 6 2010
  • Firstpage
    3953
  • Lastpage
    3955
  • Abstract
    Developments in room temperature cadmium telluride (CdTe) based solid state imaging arrays for energy-resolved single photon counting in x-ray imaging are discussed. We have designed and constructed a CdTe imaging array, 3 mm thick with a 16 × 16 grid of electrical contacts interconnected to a 256 channel ASIC, with a counting range up to 2 × 107 counts per second per mm2. Cycling between high and low flux generates measurements which are reproducible within the limits of counting statistics, with a response time less than 1 ms. Long-term stability is determined by measuring flood images repeatedly over time. Long exposure to uniform high or low flux does not affect the temporal response, indicating that no polarization is detected within the statistical error of the measurements.
  • Keywords
    X-ray imaging; application specific integrated circuits; nuclear electronics; photon counting; semiconductor counters; ASIC; CdTe detectors; electrical contacts; flood images measurement; high-flux X-ray imaging; single photon-counting; solid state imaging arrays; Biomedical imaging; Computed tomography; Detectors; Energy resolution; Photonics; Pixel; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2010 IEEE
  • Conference_Location
    Knoxville, TN
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-9106-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2010.5874556
  • Filename
    5874556