• DocumentCode
    3548590
  • Title

    Pixelated CdZnTe drift detectors

  • Author

    Kuvvetli, I. ; Budtz-Jørgensen, C.

  • Volume
    7
  • fYear
    2004
  • fDate
    16-22 Oct. 2004
  • Firstpage
    4360
  • Lastpage
    4364
  • Abstract
    A technique - the so-called drift strip method (DSM) - for improving the CdZnTe detector energy response to hard X-ray and gamma ray was applied as a pixel geometry. First tests have confirmed that this detector type provides excellent energy resolution and imaging performance. We specifically report on the performance of 3 mm thick prototype CZT drift pixel detectors fabricated using material from eV-products. We discuss issues associated with detector module performance. Characterization results obtained from several prototype drift pixel detectors are presented. Results of position and energy resolution measurements, charge sharing, IV-measurements are discussed
  • Keywords
    X-ray detection; gamma-ray detection; position sensitive particle detectors; semiconductor counters; 3 mm; drift strip method; energy resolution; gamma ray detector; hard X-ray detector; pixel geometry; pixelated CdZnTe drift detectors; prototype drift pixel detectors; room temperature semiconductor detectors; Energy resolution; Gamma ray detection; Gamma ray detectors; Geometry; Prototypes; Strips; Testing; X-ray detection; X-ray detectors; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2004 IEEE
  • Conference_Location
    Rome
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8700-7
  • Electronic_ISBN
    1082-3654
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2004.1466852
  • Filename
    1466852