• DocumentCode
    3548621
  • Title

    Quantitative evaluation of mercuric iodide and selenium for X-ray imaging device

  • Author

    Cha, Byung-Youl ; Park, Ji-Koon ; Kang, Sang-Sik ; Shin, Jung-wook ; Kim, Jin-Young ; Lee, Hyung-Won ; Nam, Sang-hee

  • Author_Institution
    Dept. of Biomed. Eng., Inje Univ., Kimhae, South Korea
  • Volume
    7
  • fYear
    2004
  • fDate
    16-22 Oct. 2004
  • Firstpage
    4496
  • Abstract
    Analog film/screen systems have been replaced with digital X-ray imaging devices using direct conversion materials. In this paper, mercuric iodide (HgI2) and amorphous selenium (a-Se) films were deposited through the particle-in-binder (PIB) and physical vapor deposition (PVD) methods, respectively. Using the MCNP 4C code, the interaction of X-ray photons in HgI2 and a-Se bulk, their transport, and transmitted energy spectrum of continuous X-ray, with total absorbed energy were simulated. Using I-V measurements, their electrical properties, such as leakage current, X-ray sensitivity, and signal-to-noise ratio (SNR), were investigated. The results of our study can be useful in the future design and optimization of direct active-matrix flat-panel detectors (AMFPD) for various digital X-ray imaging modalities.
  • Keywords
    X-ray imaging; electrical conductivity; leakage currents; mercury compounds; selenium; semiconductor counters; semiconductor materials; semiconductor thin films; vapour deposited coatings; HgI2; I-V measurements; MCNP 4C code; PVD methods; Se; X-ray photon interaction; X-ray sensitivity; active-matrix flat-panel detectors; amorphous selenium films; analog film systems; analog screen systems; continuous X-ray transmitted energy spectrum; conversion materials; digital X-ray imaging device; electrical properties; leakage current; mercuric iodide; particle-in-binder; physical vapor deposition methods; signal-to-noise ratio; Active matrix technology; Amorphous materials; Atherosclerosis; Chemical vapor deposition; Current measurement; Design optimization; Electric variables measurement; Leakage current; Signal to noise ratio; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2004 IEEE
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8700-7
  • Electronic_ISBN
    1082-3654
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2004.1466882
  • Filename
    1466882