DocumentCode :
3555396
Title :
A fully automatic, accurate method for the performance evaluation and calibration of absorber-lined chambers for radiated-emission tests
Author :
Garn, H.F. ; Müllner, Wolfgang
Author_Institution :
Austrian Res. Centre Seibersdorf, Austria
fYear :
1991
fDate :
12-16 Aug 1991
Firstpage :
19
Lastpage :
24
Abstract :
Absorber-lined chambers which do not meet the ±4 dB site acceptability criterion based on normalized site attenuation (NSA) measurements require correction factors for radiated-emission tests. Such correction factors cannot be derived from deviations between measured NSA and standard reference data because the radiation characteristics of actual equipment under test have to be taken into account. The proposed chamber calibration method provides accurate correction factors, which are valid for all typical radiated-emission test situations. These factors allow the chambers to be used for standard radiated-emission tests
Keywords :
calibration; electromagnetic compatibility; test facilities; EMC testing; absorber-lined chambers; chamber calibration method; correction factors; normalized site attenuation; radiated-emission tests; site acceptability criterion; Attenuation measurement; Automatic logic units; Calibration; Electromagnetic compatibility; Frequency; Impedance; Measurement standards; Reflection; Reflectivity; Testing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Electromagnetic Compatibility, 1991. Symposium Record., IEEE 1991 International Symposium on
Conference_Location :
Cherry Hill, NJ
Print_ISBN :
0-7803-0158-7
Type :
conf
DOI :
10.1109/ISEMC.1991.148174
Filename :
148174
Link To Document :
بازگشت