DocumentCode
3558484
Title
Adaptive filtering of the electromyographic signal for prosthetic control and force estimation
Author
Park, Euljoon ; Meek, Sanford G.
Author_Institution
Pacesetter Inc., Sylmar, CA, USA
Volume
42
Issue
10
fYear
1995
Firstpage
1048
Lastpage
1052
Abstract
An adaptive time constant filter is derived for electromyographic (EMG) signal processing in prosthetic control applications. The analysis indicates that the mean-squared estimation error can be reduced by varying the time constant of the filter as a function of the signal and its derivative. Results of several experiments indicated this filter provides faster response and smaller estimation error than several previously available filters.
Keywords
adaptive filters; artificial limbs; biocontrol; biomechanics; electromyography; force measurement; medical signal processing; adaptive time constant filter; electromyographic signal adaptive filtering; estimation error; filter response; filter time constant variation; mean-squared estimation error; prosthetic control; prosthetic force estimation; Adaptive control; Adaptive filters; Adaptive signal processing; Electromyography; Estimation error; Force control; Process control; Programmable control; Prosthetics; Signal analysis; Algorithms; Arm; Artificial Limbs; Electromyography; Equipment Design; Humans; Muscle Contraction; Time Factors;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/10.464381
Filename
464381
Link To Document